CIC Hard Drive Diagnostics Utility (Paid Version)

Supports the following Serial Numbered Equipment

  1. CIC ProTM  - Bedrock - JA
  2. CIC ProTM  - Bedrock - SCH
  3. CIC ProTM  - MP100 - SDY
  4. CARESCAPE Central Station V1 - MP100 - SGJ 
  5. CARESCAPE Central Station V2 - SKN
  6. CTS - MP100 - SEE

Usage

  1. Bedrock (Versions 4x & 5x)
    1. Insert USB Drive into an open USB slot on running Bedrock CIC.
    2. From the CIC Setup screen, select the Service Password tab.
    3. At the C:\Program Files\Marquette\CIC\4.0.x> prompt, type  “F:” and Enter
    4. In the Command prompt window type “1” and Enter.
    5. This will open a new window and then type “2” and Enter. This will launch the program.
    6. Select the Drive D: (Hard Drive) from the buttons under the Menu at the top.
      1.  NOTE: Drive C: and E: are the Compact Flash.
    7. Verify the Health Status
    8. Record the power on Hours in the spreadsheet if applicable.
  2. MP100 (Versions 5x & 6x)
    1. Insert USB Drive into an open USB slot on running MP100.
    2. From the CIC Setup screen, select the Service Password tab.
    3. At the C:\Program Files\Marquette\CIC\5.0.x> prompt, type  “H:” and Enter
    4. In the Command prompt window type “1” and Enter.
    5. This will open a new window and then type “2” and Enter. This will launch the program.
    6. Select the Drive G: from the buttons under the Menu at the top.
    7. Verify the Health Status
  3. CARESCAPE Central Station V2
    1. Insert USB Drive into an open USB slot on running MP100.
    2. From the CIC Setup screen, select the Service Password tab.​
    3. At the C:\Program Files\Marquette\CIC\5.0.x> prompt, type  “H:” and Enter
    4. In the Command prompt window type “1” and Enter.
    5. This will open a new window and then type “2” and Enter. This will launch the program.
    6. Select the Drive G: from the buttons under the Menu at the top.
    7. Verify the Health Status
  4. CTS
    1. Insert USB Drive into an open USB slot on running MP100.
    2. From the CIC Setup screen, select the Service Password tab.​
    3. At the C:\Program Files\Marquette\CIC\5.0.x> prompt, type  “H:” and Enter
    4. In the Command prompt window type “1” and Enter.
    5. This will open a new window and then type “2” and Enter. This will launch the program.
    6. Select the Drive G: from the buttons under the Menu at the top.
    7. Verify the Health Status
Screen Shots

 

CTS Hard Drive Diagnostics

 

 CIC PRO / Bedrock

 

CarescapeTM MP100 / MP100R

 

HD Spreadsheet

Data Entry Section

 

 Hard Drive by Hours